Troubleshooting
7-62
Maintenance, Troubleshooting, and Repair
Recording/
Printing
Diagnostics
The
Recording/Printing Diagnostics
tool tests the connectivity of LaserJet printers and
Philips Recorders connected to the Information Center or Client. It requires a connected
printer or recorder and a source of patient data. Clicking
Recording/Printing Diagnostics
in the
Diagnostics
menu brings up the
Printer & Recorder Diagnostics
window of
Figure 7-27 Printer & Recorder Diagnostics Window
The
Printer & Recorder Diagnostics
window has two test options --
Recording Test
and
Printing Test
.
Recording Test
tests Philips Recorders.
Recording type
provides two data source options --
Live
(an actively monitored
patient) and
Test
(an internal test recording).
Live
activates the
Available Beds
field from which a live recording from a
specific bed can be selected. A 20 second delayed recording of the primary and
secondary waves from the patient in that bed serves as the recording test.
Test
selects internal test data for the recording test.
Start
initiates the selected test recordings. The
Recording Test
output is sent to one
of the recorders in the Recorder Rack.
Stop
terminates a test in process.
Continue
produces a continuous recording.
Status from last job
indicates the status of the test (e.g.
active, done, error
).
Note
If there is more than one recorder in the Recorder Rack and the Recording Test fails, repeat
the test with only one recorder in the Rack at a time to determine which recorder is failing the
test.
Note
This window does not provide diagnostics for the 4 Channel Recorder.
Summary of Contents for IntelliVue Series
Page 152: ...IntelliVue Information Center Application Software 3 22 Software Description ...
Page 442: ...Troubleshooting 7 156 Maintenance Troubleshooting and Repair ...
Page 488: ...Procedure B 12 Web Installation on the Database Server ...
Page 492: ...Procedure C 4 External Modem Installation ...
Page 500: ...Using RAS D 8 Dial In Procedure for Remote Access to Information Center Systems ...