IntelliVue Information Center Application Software
Software Description
3-15
Sector Setup
allows the user to change the monitoring equipment associated
with a bed.
Stored Waves
sets the waves that will be stored for an individual patient.
Standby
allows temporary suspension of monitoring for a patient and the
display of standby messages in their
Patient Sector.
Alarm Limits
provides
Arrhythmia Alarm
and
ST Alarm
limits that control
conditions for triggering and turning on and off certain alarms for individual
patients.
Record/Store Alarms
controls which alarms are recorded and/or stored for
each patient. This control also affects the
Alarm Review
application that
provides review recordings of stored alarms.
ST Setup
turns ST segment measurements on and off and sets the
measurement point for an individual patient.
Telemetry Setup
sets telemetry controls for an individual patient.
Review Applications
Review Applications
provide functionality for the
Patient Data Review
buttons of
the
All Controls
window. Each review application obtains its specific data for an
individual patient from Database Services and presents it in the
Patient Window
display or delivers it to the Philips recorder or printer when requested.
Review Applications include the following for each patient.
Trend Review
provides a history of groupings of physiological parameters
plotted over time.
Alarm Review
provides a history of alarms, including waves and alarm text.
Event Review
provides a history of events detected for a patient along with
waveforms stored at the time of the event.
Wave Review
provides a history of full disclosure waveform data that have
been continuously stored.
ST Review
displays a history of acquired and derived ST waveforms and
segment values for telemetry monitored patients.
Support Applications
Support Applications
is principally software for the
Service
applications. Most of
these applications run in non-monitoring mode when patient monitoring is turned
off. They used for managing the installation, configuration, and service aspects of the
Information Center system. To access the support applications, a Service password is
required to prevent accidental termination of monitoring and unauthorized access to
system management applications.
Support software contains the following features.
•
Support Logs
•
Diagnostic Tools
Summary of Contents for IntelliVue Series
Page 152: ...IntelliVue Information Center Application Software 3 22 Software Description ...
Page 442: ...Troubleshooting 7 156 Maintenance Troubleshooting and Repair ...
Page 488: ...Procedure B 12 Web Installation on the Database Server ...
Page 492: ...Procedure C 4 External Modem Installation ...
Page 500: ...Using RAS D 8 Dial In Procedure for Remote Access to Information Center Systems ...