Configuring the System
Start-up and Configuration
8-7
•
When sighting the measured edges of the
Main Screen
, close one eye and use a line-
of-sight that is perpendicular to the ruler.
Note
To obtain a 25 mm/s sweep speed on the Second display, the Primary and Second displays
need to be the same size.
Step 5.
Repeat
Steps 1-4
for all
CRT Displays
of the Philips system.
For Flat Panel
Displays
Step 1.
Press the
PROCEED
button on the front panel of the display to bring up the
On-
Screen Manager
window.
Note
If the
LOCK OUT
mode is displayed, hold down the
PROCEED
button and simultaneously
press and release the
CONTROL
button. Then release the
PROCEED
button.
If the
On-Screen Manager
appears but has the
wrong orientation
, press and release the
display
EXIT
button and then press the display
RESET
button.
Step 2.
Press and release the
CONTROL
button to highlight
Auto
.
Step 3.
Press and release the
PROCEED
button.
The display will briefly go blank and then reappear. This step automatically adjusts the image
position, horizontal size, and the fine display settings. The borders shall now be centered and
fill the screen.
Note
To reactivate the
ON-Screen Manager
LOCKOUT
mode:
•
Press
EXIT
twice to activate the
On-Screen Manager
.
•
Hold down the
PROCEED
button and simultaneously press and release the
CONTROL
button.
•
Immediately press and release the
CONTROL
button.
•
Release the
PROCEED
button.
•
Skip
Step 4
.
Step 4.
Press and release the
EXIT
button.
Step 5.
Repeat
Steps 1-4
for all
Flat Panel Displays
of the system.
Verifying
Sweep Speed
Accuracy
Step 1.
Verify the accuracy of the 25mm/s setting as follows:
Summary of Contents for IntelliVue Series
Page 152: ...IntelliVue Information Center Application Software 3 22 Software Description ...
Page 442: ...Troubleshooting 7 156 Maintenance Troubleshooting and Repair ...
Page 488: ...Procedure B 12 Web Installation on the Database Server ...
Page 492: ...Procedure C 4 External Modem Installation ...
Page 500: ...Using RAS D 8 Dial In Procedure for Remote Access to Information Center Systems ...