Configuring the System
8-14
Start-up and Configuration
Table 8-4. M3170 - Test and Inspection Matrix
Test Block
Name
Test or Inspection to Perform
Expected Results
What to Record
on Service
Record
Visual
For installation, perform visual inspection of shipping
cartons and contents.
No visible damage
Expected answer is
“Yes”. If so,
Visual
test
passed.
V:P
or
V:F
where
P
= Pass
F
= Fail
Power On:
Power on all devices.
Observe software boot up.
Observe that support display shows that the DataServer
Startup was complete.
Devices power up into
expected status on
support display
PO:P
or
PO:F
where
P
= Pass
F
= Fail
Performance:
Verify purchased options are enabled.
Verify normal system operation.
Do LAN based monitors record and print requests respond
to user inputs?
Expected answers are
“Yes”.
If so,
Performance
test
passed.
P:P
or
P:F
where
P
= Pass
F
= Fail
If #C14 HL7 Export is installed, ping by host name the
target HL7 client device
Expected answer is
“Yes”. If so,
Performance
test passed.
PHL7:P
or
PHL7:F
where
P
= Pass
F
= Fail
Safety
Is the device connected to a 78581A/B (SCC) or telemetry
mainframe?
If answer is “No”, no
Safety
test required.
If answer is “Yes”,
move to next step.
S:NA
where
NA
= Not
required
Is there a 2nd earth ground connected to the SCC or
telemetry mainframe
Expected answer is
“Yes”.
If so,
Safety
test passed.
S:P
or
S:F
where
P
= Pass
F
= Fail
Confirm that 78581A/B or telemetry mainframe has
passed Safety Test Block (see
78581A/B
or
Telemetry
Mainframe
Test and Inspection Matrix
).
If answer is “Yes”,
Safety
test passed.
If answer is “No”,
move to next step.
S:P
or
S:F
where
P
= Pass
F
= Fail
Perform
78581A/B
or
Telemetry Mainframe Safety
Test Block
Expected answer is
“Yes”. If so,
Safety
test
passed
S:P
or
S:F
where
P
= Pass
F
= Fail
Summary of Contents for IntelliVue Series
Page 152: ...IntelliVue Information Center Application Software 3 22 Software Description ...
Page 442: ...Troubleshooting 7 156 Maintenance Troubleshooting and Repair ...
Page 488: ...Procedure B 12 Web Installation on the Database Server ...
Page 492: ...Procedure C 4 External Modem Installation ...
Page 500: ...Using RAS D 8 Dial In Procedure for Remote Access to Information Center Systems ...