Troubleshooting
7-32
Maintenance, Troubleshooting, and Repair
RecMgr
Error: 2515, One of the waves needed for the recording is no
longer available.
If this occurs around a patient
discharge or transfer or when a
bedside monitor module is pulled
out, no action required.
If occurs at other times, reboot system.
Error: 2516, The requested memory could not be allocated.
Reboot system. If this persists, reload
Philips software.
Error: 2517, Data from the pds has not arrived yet. Indicates a
long time delay in retrieving data from the pds.
If this occurs around a patient
discharge or transfer or when a
bedside monitor module is pulled
out, no action required.
If occurs at other times, reboot system.
Error: 2518, One or more threads in this process failed to start.
Problem is likely due to a system wide failure and may require
a reboot of the machine.
Reboot system. If this persists, reload
Philips software.
Error 250C, The recorder corba id key was missing or could
not be read or written to.
Reinstall Philips software.
Error: 250D, A corba call failed
Reboot system. If this persists, reinstall
Philips software
Error: 250E, Programmer error. Please submit a bug report
with the text from this message.
Report text in
Event Log
to Response
Center.
Reboot system.
SDNConfig
The database service has not started. The application must be
terminated
Reboot system.
sdprocess
Clinical password
entered
Clinical password was used in
Unit
Settings
password window
Verify proper user access
ServiceLog
EntryApp
Winsock connection failed
Reboot system. If this persists, reinstall OS
software
Could not initialize and/or update the application. The
application was terminated
Reinstall Philips software
All devices on the network are unavailable. Application
terminated.
Check SDN cable and connections, test
SCC
Cannot connect to database. Application terminated.
Reboot system. If this persists, reinstall
Philips software
EMS Service Log Database is corrupt.
Shut down the
Service Log
and Contact
the Response Center
Application could not be started.
Reboot system. If this persists, reinstall
Philips software
ServiceLog
ViewerApp
Winsock connection failed
Reboot system. If this persists, reinstall OS
software
Could not initialize and/or update the application. The
application was terminated
Reinstall Philips software
All devices on the network are unavailable. Application
terminated.
Check SDN cable and connections. Test
SCC
Cannot connect to database. Application terminated.
If this persists, reinstall Philips software
Table 7-7. Application Events
Source
Description
(Event Code)
Explanation
Corrective Action
Summary of Contents for IntelliVue Series
Page 152: ...IntelliVue Information Center Application Software 3 22 Software Description ...
Page 442: ...Troubleshooting 7 156 Maintenance Troubleshooting and Repair ...
Page 488: ...Procedure B 12 Web Installation on the Database Server ...
Page 492: ...Procedure C 4 External Modem Installation ...
Page 500: ...Using RAS D 8 Dial In Procedure for Remote Access to Information Center Systems ...