Overview
5-2
Hardware Installation
The noise immunity of fiber optic cable is superior to UTP cable so that fiber optic cable
should be used for any 10 Mbit/s cable runs over 100 m for which RF or electrical noise is
a potential problem.
UTP Cable Plant
Installation
A typical cable plant installation for UTP Category 5 cables for Information Centers,
Clients and Clinical Network/Database Server components is shown in
.
Figure 5-1 Typical UTP CAT5 Cable Plant Installation
The UTP CAT5 cable plant should meet the following:
•
Patch panels for all switches
should be in the wiring closets where switches will be
installed
•
RJ-45 wall boxes or patch panels for repeaters and extension switches
should be
in closets where they will be installed. Repeaters and Extension Switches should
not
be located above a ceiling.
•
RJ-45 Wall boxes for Information Centers, Clients, Printers, and Server,
should
be within patch cable lengths of their devices.
•
Cabling, patch panels, switches, repeaters, and media translators
should be more
than 1 m (3 ft.) from all powered devices (Server, UPS, etc.).
•
Labels on all UTP CAT5 cables and terminations
should identify the cable, patch
panel, port number, and wall box termination.
•
Test Documentation
should demonstrate that the UTP CAT5 cable plant meets
CAT5 standards for NEXT, attenuation, wiremap, and length.
Caution
In-wall cabling - UTP and fiber optic - must be terminated at a patch panel or wall
box and not directly at an active Network device.
Wiring Closet for DBS
Wiring Closet for Repeater
Test Documentation
Patch Panel
CATEGORY 5
CATEGORY 5
262-1
262-1
CATEGORY 5
CATEGORY 5
262-1
262-1
CATEGORY 5
CATEGORY 5
262-1
262-1
CATEGORY 5
CATEGORY 5
262-1
262-1
CATEGORY 5
CATEGORY 5
262-1
262-1
1
2
3
4
5
6
7
8
9
10 11 12
13 14 15 16 17 18
19 20 21 22 23 24
AMP
Information
Center
Client
Information
Center
Printer
Summary of Contents for IntelliVue Series
Page 152: ...IntelliVue Information Center Application Software 3 22 Software Description ...
Page 442: ...Troubleshooting 7 156 Maintenance Troubleshooting and Repair ...
Page 488: ...Procedure B 12 Web Installation on the Database Server ...
Page 492: ...Procedure C 4 External Modem Installation ...
Page 500: ...Using RAS D 8 Dial In Procedure for Remote Access to Information Center Systems ...