Clinical Applications
Introducing the IntelliVue Information Center System
1-35
ST Segment
Monitoring
The Information Center system also provides ST segment analysis for adult patients
monitored via telemetry only. ST segments can be analyzed for up to six leads of ECG
monitoring, depending on the type of telemetry monitor and patient cable used, and for both
non-paced and atrially paced patients (but not ventricularly paced patients).
Note
ST segment analysis for patients monitored by bedside monitors is performed at the bedside
monitor and is not available on the Information Center.
Details of displayed ST data, ST alarms, and user interfaces are provided in the
IntelliVue
Information Center Instructions for Use
. Details of the ST Segment algorithm are
described in the
ST Segment Monitoring Application Note
provided on the Documentation
CDROM.
EASI 12-Lead
ECG Review
and Report
If the bedside or telemetry system has the EASI 12-Lead capability (a method of deriving 12
ECG Leads using 5 electrodes), the clinician can view all available leads from the Patient
Window - see table below. In addition, the clinician can request a 12-Lead report. Details of
EASI 12-Lead ECG monitoring are provided in the Information Instruction for Use.
Patient Data
Review
The Information Center system has extensive patient data review capability.
Patient Data
Review
windows display patient physiological parameters and alarm events that have been
collected over time from bedside and telemetry monitors. Review windows can be designed in
a variety of formats to facilitate clinician evaluations of patient status and decisions of
diagnosis, prognosis, medication, and patient transfer and discharge.
The following review applications are accessible from the
Patient Data Review
buttons in
the
All Controls
window.
Alarm Review
Alarm Review
shows the alarm events that have been automatically stored during patient
monitoring and alarm strips that were saved for each alarm. A typical
Alarm Review
record
for a single display is shown in
. Stored alarms are those that were configured in
Alarm Management and Setup
and have been stored in the alarm history database during
patient monitoring. Up to 50, 30-second alarms can be stored in the standard system and up to
150, 30-second alarms as a purchased option. Each stored alarm includes the date and time,
alarm text, and a 30-second compressed waveform. Systems with a single display can display
5 alarm strips, and systems with dual display, 10 alarm strips.
Saved strips are those that are manually saved from
Wave Review
,
Event Review
, or
Alarm
Review
windows. These strips are labeled “
Saved
” and can be viewed along with stored
alarms. Up to 10 strips can be saved in the standard system and up to 30 for the 150, 30-
# Cable
Wires
Philips
Telemetry
Digital UHF
Telemetry
3
1 lead
1 lead
4
up to 6 leads
5
up to 6 leads
2 leads
Summary of Contents for IntelliVue Series
Page 152: ...IntelliVue Information Center Application Software 3 22 Software Description ...
Page 442: ...Troubleshooting 7 156 Maintenance Troubleshooting and Repair ...
Page 488: ...Procedure B 12 Web Installation on the Database Server ...
Page 492: ...Procedure C 4 External Modem Installation ...
Page 500: ...Using RAS D 8 Dial In Procedure for Remote Access to Information Center Systems ...