Configuring the System
Start-up and Configuration
8-5
Power
Failure
Response
Test
This test verifies the operation of the
UPS
in providing power in case of a loss of power to the
Philips system.
Step 1.
Disconnect the power cord of the UPS from the wall outlet or Power Distribution
Module for about
3 seconds
.
Step 2.
Verify that the Philips system continues to operate and the UPS gives and audible
tone.
Step 3.
Restore the power cord connection.
Step 4.
From the
All Controls
page, select the
Service
button and enter the service password.
Step 5.
Select
Support Logs -> Event Logs-> System
Step 6.
Review the log and verify that there are two UPS entries at the time of the test: one
indicating a power failure was detected, and one indicating that power was restored.
Setting
Display
Sweep
Speed
For full compliance with Philips system specifications, the sweep speed of the display(s) shall
be 25mm/s + 10%. Setting this value is required for any new or replaced display, but not for
replacement of the video card or video driver.
The following procedures describe how to make the sweep speed setting. The procedure for
CRT displays is different from that for Flat Panel Displays.
For CRT
Displays:
Step 1.
Set the image on the
Primary Display
to the
Main Screen
.
Step 2.
Access the
Image Controls
of the Primary Display located in the lower section of the
front panel.
Step 3.
Using the
horizontal and vertical image position controls
, position the
Main
Screen
image as close to the exact center of the display screen as possible. Center the
image on the CRT screen by eye by equalizing the size of the black boundaries between
the image and the plastic frame. Consult the User’s Manual for your display if you are
unsure which controls to use.
Equal
Equal
Summary of Contents for IntelliVue Series
Page 152: ...IntelliVue Information Center Application Software 3 22 Software Description ...
Page 442: ...Troubleshooting 7 156 Maintenance Troubleshooting and Repair ...
Page 488: ...Procedure B 12 Web Installation on the Database Server ...
Page 492: ...Procedure C 4 External Modem Installation ...
Page 500: ...Using RAS D 8 Dial In Procedure for Remote Access to Information Center Systems ...