Locating System Components
5-14
Hardware Installation
M3169 Small
Database
Server
Installation
The M3169 Small Database Server should be installed in its intended location. This includes
the Server, display, keyboard, and, mouse. The display, keyboard, and mouse are intended for
a work surface.
Step 1. Install the Small Database Server
components in their intended positions.
Step 2. Install the 650 VA UPS
with the proper voltage for the Server as follows:
–
Turn
Off the UPS
On/Off switches.
–
Connect the battery wire
of the 650 VA UPS.
–
Connect the UPS input power cord
to a properly grounded electrical output
–
Verify PC Voltage select switch setting
–
Connect the
PC power cord
to a UPS battery backup outlet.
–
Verify the UPS Switch settings are accurate
–
Connect the Philips 9 pin UPS communication cable (gray)
to the computer
interface port of the UPS and the Serial A port on the rear of the Server. See
Figure 5-14 Connections to the M3169 Small Database Server D510 PC
OVERLOAD
PROCTECTOR
UPS OUTPUT
120V 50/60 Hz
INPUT
TOTAL
4 3 2 1
COMPUTER
INTERFACE PORT
SITE
WARNING
FAULT
ACCESSORY
SURGE
PROTECTION
BATTERY
BACKUP
OUTPUT
CURRENT
12A MAX
120V-
50-60 Hz
12A
Power Cord Connector
Voltage Select
Mouse Connector
Keyboard Connector
Hospital LAN
Display Connector
Philips 9 Pin UPS/Server
Communication
Cable (gray)
Switch
LAN Connector
10BT
100TX
-
Card (Optional)
Summary of Contents for IntelliVue Series
Page 152: ...IntelliVue Information Center Application Software 3 22 Software Description ...
Page 442: ...Troubleshooting 7 156 Maintenance Troubleshooting and Repair ...
Page 488: ...Procedure B 12 Web Installation on the Database Server ...
Page 492: ...Procedure C 4 External Modem Installation ...
Page 500: ...Using RAS D 8 Dial In Procedure for Remote Access to Information Center Systems ...