Troubleshooting
Maintenance, Troubleshooting, and Repair
7-29
Table 7-7. Application Events
Source
Description
(Event Code)
Explanation
Corrective Action
dataserver
Error: 2000
Monitoring conflict or Central Station
# conflict. A bed is being monitored by
more than one Information Center
Correct SDN branch assignments in
SDN
Config
and
Display Setup
to assure that
each monitor is assigned to only one
Central Station
Error: 2001
Patient Data Store told DataAcquisition
to tune more than 16 beds
simultaneously
Contact Response Center
Error: 2002
Network error. SCC probably
disconnected or turned off.
- Check SCC cable and SDN interface
connection
- Check that SCC is turned on.
Error: 2004
Length of message block received from
Redirector does not equal the first word
of the buffer
Contact Response Center
Error: 2005
Invalid link to next SDN message.
Corrupt incoming data.
Contact Response Center
Error: 2006
SDN message too small
Contact Response Center
Error: 2007
HRC check failed. Corrupt incoming
data
Contact Response Center
Error: 2008
Invalid link to next SDN message.
Corrupt incoming data.
Contact Response Center
Error: 2009
Invalid high order bits in SDN
message. Corrupt incoming data.
Contact Response Center
Error: 200B
Virtual Terminal Protocol could not
connect to arrhythmia controls object.
Contact Response Center
Error: 200C
Virtual Terminal Protocol could not
connect to database alarms object
Contact Response Center
Error: 200D
Virtual Terminal Protocol error in
GetAlarmWave
Contact Response Center
Error: 200E, 200F,
2010, 2011, 2106,
2017, 2018, 2019
Low level Data Control Protocol error
No action required
Error: 201A
Data Acquisition Receive thread is
shutting down
Not an error, no action required
Error: 201B
Low level Data Control Protocol error
No action required
Error: 2021
Branch number changed. User altered
SCC configuration while system was
running
- Check SCC cabling
- Shut down Information Center and restart.
System should recognize new
branch and start normally.
Error: 2022
SDN Interface Card reset
Shut down Information Center and restart
Error: 2023
Data Link Control header incorrectly
terminated
Reseat or replace SDN Interface Card
Error: 2024
SDN Data block incorrectly terminated Reseat or replace SDN Interface Card
Summary of Contents for IntelliVue Series
Page 152: ...IntelliVue Information Center Application Software 3 22 Software Description ...
Page 442: ...Troubleshooting 7 156 Maintenance Troubleshooting and Repair ...
Page 488: ...Procedure B 12 Web Installation on the Database Server ...
Page 492: ...Procedure C 4 External Modem Installation ...
Page 500: ...Using RAS D 8 Dial In Procedure for Remote Access to Information Center Systems ...