Troubleshooting
7-114
Maintenance, Troubleshooting, and Repair
Figure 7-63 UPS Configuration Window
Step 5.
Verify (or enter) the following information in the fields of the
UPS
configuration
window.
Uninterruptible Power Supply is installed on:
COM1:
UPS Configuration
UPS Interface Voltages:
Power failure signal
Positive
Low battery signal at least
2 minutes before shutdown
(Disabled)
Remote UPS Shutdown
Positive
Execute Command File
(Disabled)
UPS Characteristics
Expected battery Life:
2 min.
Battery Recharge time
per minute of run time:
1 min.
UPS Service
Time between power failure
and initial warning message:
120 sec.
Delay between warning
messages:
120 sec.
Step 6.
When the UPS configuration settings have been verified, click
OK
to return to the
Windows Main Menu
.