![background image](http://html1.mh-extra.com/html/omron/fq2-s4/fq2-s4_user-manual_742879518.webp)
External Reference Parameters
516
FQ2-S4 User’s Manual
2D Codes (DPM)
External
refer-
ence
number
Category Data name
Set/acquired
Data range
Default
Text string
assigned
to
Ethernet
output
setting
Logged
0
Mea-
sure-
ment
result
Judgment
Acquired only -2: No judgment (not
inspected)
0: Judgment result OK
-1: Judgment result NG
-13: Teaching not executed
error
-14: Figure not registered
error
-15: Outside region error
-16: Inspection timeout
error
-2
JG
Logged
5
Num. of char.
Acquired only 0 to 1024
0
N
Logged
6
*1
Characters
Acquired only Text string
---
---
---
8
Cell Recog. Rate
Acquired only 0 to 100
0
E
Logged
9
Contrast
Acquired only 0 to 100
0
C
Logged
10
Focus
Acquired only 0 to 100
0
F
---
17
Index number
Acquired only -2: Verification OFF or read-
ing error, -1: Verification is
NG, 0 to 31: Master data
No.
−
2
IN
Logged
103
Output
parame-
ter
Reflect
Setting/Acqui-
sition
0: Yes, 1: No
0
---
---
Summary of Contents for FQ2-S4
Page 1: ...User s Manual Smart Camera FQ2 S4 Cat No Z330 E1 01 ...
Page 58: ...Setting Up Ethernet 56 FQ2 S4 User s Manual MEMO ...
Page 214: ...Calculations and Judgements Using Inspection Item Data 212 FQ2 S4 User s Manual MEMO ...
Page 234: ...Adjusting Judgement Parameters during Operation 232 FQ2 S4 User s Manual MEMO ...
Page 458: ...Basic Troubleshooting 456 FQ2 S4 User s Manual MEMO ...
Page 591: ...Index FQ2 S4 User s Manual 589 Index 12 ...
Page 593: ......