Reading Bar Codes
FQ2-S4 User’s Manual
4
Setting Up
Inspect
ions
119
Measurement Data That Can Be Logged (Bar Code)
The following values can be logged as measurement data.
* When logging data is output, the data is output in the order of the above table.
7-6 Logging Measurement Data and Image Data: p. 252
Errors
Errors in Teaching
A teaching error message is displayed If bar codes cannot be read during an automatic registration.
The reading may be unstable due to low contrast.Adjust the brightness to improve the contrast of the bar code.
Adjusting the Brightness: p. 61
Measured item
Range of value
Description
Judgement
−
2: No judgement (not measured),
0: Judgement is OK,
−
1: Judgement is NG,
−
13: Teaching not performed error,
−
14: Figure not registered error,
−
15: Out of range error,
−
16: Measurement timeout error
This is the judgement result from the measurements.
Num. of characters
0 to 1024
The number of characters read is output.