Maintenance and Instrument Interfaces
R&S
®
UPP
882
User Manual 1411.1055.32 ─ 10
Table 9-2: Universal commands
Command
VISUAL BASIC command
Effect on instrument
DCL (Device Clear)
IBCMD (controller%, CHR$(20))
Terminates processing of the
commands that have just been
received, and sets the command-
processing software to a defined
initial state. Does not modify the
instrument configuration.
IFC (Interface Clear)
IBSIC (controller%)
Resets the interfaces to their ini-
tial setting.
LLO (Local Lockout)
IBCMD (controller%, CHR$(17))
The manual LOCAL switchover
function is disabled.
SPE (Serial Poll Enable)
IBCMD (controller%, CHR$(24))
Ready for serial polling.
SPD (Serial Poll Disable)
IBCMD (controller%, CHR$(25))
End of serial polling.
PPU (Par. Poll Unconfigure)
IBCMD (controller%, CHR$(21))
End of parallel polling query sta-
tus.
Addressed commands
The addressed commands are in the code range 0x00 to 0x0F. They only act on instru-
ments which are addressed as listeners.
Table 9-3: Addressed commands
Command
VISUAL BASIC command
Effect on instrument
SDC (Selected Device Clear)
IBCLR (device%)
Terminates processing of the
commands that have just been
received, and sets the command-
processing software to a defined
initial state. Does not modify the
instrument configuration.
GET (Group Execute Trigger)
IBTRG (device%)
Triggers a measurement or a
sweep.
GTL (Go to Local)
IBLOC (device%)
Transition to the "Local" state
(manual operation).
PPC (Parallel Poll Configure)
IBPPC (device%, data%)
Configure instrument for parallel
poll. The VISUAL BASIC com-
mand also executes PPE/PPD.
9.4.2
Ethernet Interface (LAN)
As standard, the instrument is equipped with an Ethernet interface. The connector
socket is at the rear of the instrument. Network administrators with the appropriate
rights can transfer files across the network and use network resources such as network
directories or printers. The instrument can also be remote-controlled and manually
operated in the network. A commercially available shielded RJ-45 cable is used for
connection.
The network card operates with 100-MHz-Ethernet IEEE 802.3u. The TCP/IP network
protocol and the associated network services are preconfigured. TCP/IP provides con-
Instrument Interfaces
Содержание R&S UPP
Страница 1: ...R S UPP Audio Analyzer User Manual User Manual Version 10 1411105532 ÅP ...
Страница 34: ...Contents R S UPP 18 User Manual 1411 1055 32 10 ...
Страница 630: ...Instrument Functions R S UPP 614 User Manual 1411 1055 32 10 Format ...
Страница 631: ...Instrument Functions R S UPP 615 User Manual 1411 1055 32 10 Format ...