Instrument Functions
R&S
®
UPP
365
User Manual 1411.1055.32 ─ 10
"Auto"
Where possible, the RMS measurement is performed over entire AC
periods; with medium or high frequencies, the measurement time is
increased so that the maximum algorithmic error does not exceed
0.05
%.
With very weak, low-frequency or pure DC signals, the maximum
measurement time is limited to 200
ms (corresponding to 2 periods at
10
Hz).
Recommended for single-tone measurements using an unknown
external signal source if maximum measurement accuracy is desired.
Note:
For infrasound measurements (below 10
Hz), can only be mea-
sured using a fixed measurement time (Value or Gen Track).
"Value "
The RMS measurement is performed over a preset time which must
be entered in the parameter field below.
To prevent errors due to termination of measurements, the selected
measurement time should be an integer multiple of the period. If the
frequency of the signal to be measured is unknown or not constant,
the selected measurement time should be as long as possible in
order to minimize such errors.
Recommended for
●
AC measurements of a known external signal source
●
measurements of severely noise-corrupted or distorted signals of
unknown frequency
●
measurements of noise signals or multi-tone signals
Level Measurements
Содержание R&S UPP
Страница 1: ...R S UPP Audio Analyzer User Manual User Manual Version 10 1411105532 ÅP ...
Страница 34: ...Contents R S UPP 18 User Manual 1411 1055 32 10 ...
Страница 630: ...Instrument Functions R S UPP 614 User Manual 1411 1055 32 10 Format ...
Страница 631: ...Instrument Functions R S UPP 615 User Manual 1411 1055 32 10 Format ...