Manual Operation
R&S
®
UPP
194
User Manual 1411.1055.32 ─ 10
Number in red
Measuring range has been overranged; this can occur if:
●
Range
Fix setting is being used
●
A signal with a level at range limit is applied with a crest factor
>
2
●
Common-mode voltage is applied in BAL input configuration
Number in black
Measuring range has been underranged; this can occur if Range Fix or Lowest
setting is being used
Other status display
Single
Single measurement is being performed
Cont
Continuous measurement is being performed
Terminated
Single measurement has been terminated
Stopped
Measurement has been stopped
Wait for Trigger
Analyzer is waiting for set trigger condition
Generator status
Gen Off
Generator is deactivated
Gen Running
Generator signals are being output
Output Off
All output lines are deactivated
Gen Overload
Generator has been deactivated owing to overload
Sweep status
Sweep Off
No sweep is being performed
Sweep Waiting
Sweep has been defined but not yet started
Sweep Terminated
Single sweep has been terminated
Sweep Stopped
Sweep has been stopped but can be continued
Sweep Run Cont
Continuous sweep is being performed
Sweep Run Single
Single sweep is being performed
Sweep Run Manual
Manual sweep is being performed
Overview of Status Messages
Содержание R&S UPP
Страница 1: ...R S UPP Audio Analyzer User Manual User Manual Version 10 1411105532 ÅP ...
Страница 34: ...Contents R S UPP 18 User Manual 1411 1055 32 10 ...
Страница 630: ...Instrument Functions R S UPP 614 User Manual 1411 1055 32 10 Format ...
Страница 631: ...Instrument Functions R S UPP 615 User Manual 1411 1055 32 10 Format ...