Instrument Functions
R&S
®
UPP
419
User Manual 1411.1055.32 ─ 10
Start
Start value of the FFT frequency band (display only)
Remote command:
Stop
End value of the FFT frequency band (display only)
In the case of a broadband FFT in the digital analyzer, this is 23 or 48 times the sam-
pling frequency; in the case of an undersample FFT, the value here is the "Span" value
less one bin.
Remote command:
Resolution
Displays the frequency resolution.
The frequency resolution is the quotient of sampling rate and FFT size.
Remote command:
SENSe[1]:FUNCtion:FFT:RESolution?
Meas Time
Displays the integration time.
The integration time is the quotient of FFT size and sampling rate.
Remote command:
5.27
Noise Voltage Density
Measurement of Noise Voltage Density is no extra measurement function, but is inte-
grated in the FFT Display.
Noise Voltage Density