Remote Control - Command Description
R&S
®
UPP
724
User Manual 1411.1055.32 ─ 10
Manual operation:
See
SENSe[1]:VOLTage:RANGe<n3>:VALue
<value>
Suffix:
<n3>
.
<n3> = 1 with setting
INPut:RANGe:CHANnels TRACk
.
<n3> = 1 to 8 with setting
INPut:RANGe:CHANnels SPLit
. It
denotes channel 1 to 8.
Parameters:
<value>
float
range nominal value or any voltage value
Manual operation:
See
Table 7-1: Table for analog Analyzer
value = 200 mV
(0.00 ... 0.21999 V)
value = 800 mV
(0.22 ... 0.87999 V)
value = 3 V
(0.88 ... 3.29999 V)
value = 12 V
(3.3 ... 13.19999 V)
value = 50 V
(13.2 ... 55.00 V)
The query response to this command returns the range nominal value in Volts, e.g.,
SENSe:VOLTage:RANGe2:VALue?
Response =
0.8
for the range 800 mV RMS
Any voltage value can be specified instead of the range nominal value to specifically
set the respective measurement range. This range is shown in brackets.
7.4.3
Configuring the Digital Analyzer
The settings for the digital analyzer are performed in the
INPut
and
SENSe
subsys-
tems.
●
●
●
●
●
INPut[1]:SAMPle:FREQuency:MODE
●
●
Analyzer Settings