Instrument Functions
R&S
®
UPP
473
User Manual 1411.1055.32 ─ 10
Remote command:
SENSe[1]:UFILter<n2>:PASSb:LOWer
Passb Upp
Used to specify the lower corner frequency (- 0.1
dB) of the passband for bandpass or
bandstop filters.
Remote command:
SENSe[1]:UFILter<n2>:PASSb:UPPer
Stopb Low
Displays the lower cutoff frequency of the stopband in the case of bandpass and band-
stop filters; the cutoff frequency is derived from the selected passband and the stop-
band attenuation. This value is only updated after the filter is activated.
Remote command:
SENSe[1]:UFILter<n2>:STOPb:LOWer
Stopb Upp
Displays the upper cutoff frequency of the stopband in the case of bandpass and band-
stop filters; the cutoff frequency is derived from the selected passband and the stop-
band attenuation. This value is only updated after the filter is activated.
Remote command:
SENSe[1]:UFILter<n2>:STOPb:UPPer
5.40.4
Notch Filter, Third-Octave Filter and Octave Filter
Third-octave and octave filters are bandpass filters whose passband is geometrically
symmetrical to the center frequency, whereby the passband frequencies are derived
from the center frequency as follows:
Figure 5-20: Notch filter frequency response with designation of the filter parameters
User-Definable Filters