Remote Control - Fundamentals
R&S
®
UPP
641
User Manual 1411.1055.32 ─ 10
Addresses from 1 to 30 are permissible. After setting a new address, the device status
of the R&S
UPP is saved with the new address. This prevents the R&S
UPP from „for-
getting" the new address if switched OFF with the power switch rather than being prop-
erly shut down.
Sending the first command starts remote control.
6.3.2
Remote Control via LAN Interface
For remote control through a network, the controller and the instrument must be con-
nected via the LAN interface to a common network with TCP/IP network protocol.
Connection of the R&S
UPP to a network and querying of the computer name is
Chapter 2.10, "Connect R&S UPP to a Network (LAN)"
R&S
UPP is supplied with an activated Windows firewall already installed. In contrast
to file transfer via a LAN or manual remote operation with Remote Desktop, which
requires only partial disablement (exception) in the firewall, remote control via a LAN is
possible only if the firewall is completely deactivated.
Setting remote control:
"Start" - "Programs" - "Windows Firewall" - "Off (not recommended)":
Changing to Remote Control
Содержание R&S UPP
Страница 1: ...R S UPP Audio Analyzer User Manual User Manual Version 10 1411105532 ÅP ...
Страница 34: ...Contents R S UPP 18 User Manual 1411 1055 32 10 ...
Страница 630: ...Instrument Functions R S UPP 614 User Manual 1411 1055 32 10 Format ...
Страница 631: ...Instrument Functions R S UPP 615 User Manual 1411 1055 32 10 Format ...