Instrument Functions
R&S
®
UPP
600
User Manual 1411.1055.32 ─ 10
In addition to the red dots in the trace or the red bars in the bar graph, in the lower right
corner of the graphic display and the measured value list there is an indicator of limit
violations, so that it is possible to see at a glance whether any limit violation exists. The
interpretation is as follows:
Display for limit violations
"Limit Upper"
Limit Upper is activated and there is no limit violation.
"Limit Lower"
Limit Lower is activated and there is no limit violation.
"Limit Upper
past"
The "Limit Upper" value has been violated in the past but there is cur-
rently no limit violation. The violation is indicated until the [START ]
key is pressed again.
"Limit Lower
past"
The "Limit Lower" value has been violated in the past but there is cur-
rently no limit violation. The violation is indicated until the [START ]
key is pressed again.
"Limit Upper
current"
The "Limit Upper" value is currently being violated.
"Limit Lower
current"
The "Limit Lower" value is currently being violated.
5.50.2
Measurement Results
A lower and upper limit can be specified in any measurement result configuration
panel:
In the combo display, the limits are displayed as red limit markers (brackets) in the bar
graph and as numeric values:
Limit Monitoring
Содержание R&S UPP
Страница 1: ...R S UPP Audio Analyzer User Manual User Manual Version 10 1411105532 ÅP ...
Страница 34: ...Contents R S UPP 18 User Manual 1411 1055 32 10 ...
Страница 630: ...Instrument Functions R S UPP 614 User Manual 1411 1055 32 10 Format ...
Страница 631: ...Instrument Functions R S UPP 615 User Manual 1411 1055 32 10 Format ...