Instrument Functions
R&S
®
UPP
433
User Manual 1411.1055.32 ─ 10
A maximum of 48000
samples can be stored. Apart from the bandwidth (analog) and
sampling rate (digital), the possible recording length also depends on the selected
operating mode (the compression factor):
●
The higher the bandwidth or sampling rate is, the shorter the maximum possible
recording length will be.
●
The recording length can be multiplied by using one of the compression modes.
Note:
If a multichannel analyzer is used, max. 1.44
million samples can be monitored.
This corresponds to, for example, 15
s at an analog bandwidth of 40
kHz. Therefore, if
one of the compression modes is used, the recording length is automatically limited to
this time.
The recording length does not depend on the length of the time axis (selected in the
"Waveform Config" panel).
The lower the selected recording length is, the higher the update rate for the graphic
will be.
If a long recording time is set, any section of the recorded samples can be analyzed by
reducing the graphical time axis (also possible at a later stage).
Note:
The selected "Trace Length" (not the length of the displayed trace section) deter-
mines the recording time for each measurement and thus also the total measurement
time. If a high measurement speed is required (e.g. for sweeps or remote-control
mode), the memory depth should not be longer than the measurement time of the
measurement function.
Remote command:
Trigger Chan
Used to select the channel whose event triggers a measurement.
If a measurement channel which was inactive is selected as a trigger channel, it is
automatically switched on. A warning appears if the channel selected as a trigger
channel is switched off.
Switching off the measuring channel, which was selected as the trigger channel, there-
fore changes the trigger channel after this warning has been acknowledged.
Remote command:
Trigger Source
Used to activate the trigger and to select the trigger source.
Analysis in the Time Domain
Содержание R&S UPP
Страница 1: ...R S UPP Audio Analyzer User Manual User Manual Version 10 1411105532 ÅP ...
Страница 34: ...Contents R S UPP 18 User Manual 1411 1055 32 10 ...
Страница 630: ...Instrument Functions R S UPP 614 User Manual 1411 1055 32 10 Format ...
Страница 631: ...Instrument Functions R S UPP 615 User Manual 1411 1055 32 10 Format ...