Instrument Functions
R&S
®
UPP
237
User Manual 1411.1055.32 ─ 10
Freq Ch1
Frequency of Channel 1 (Freq Mode Ch1&2).
Note:
If Gen Track is selected as the measurement time in the analyzer, measurement
channel 1 follows the frequency of generator channel 1. If necessary, the generator fre-
quency is modified slightly so that the signal periods measured during the measure-
ment time contain a sufficient number of samples.
Remote command:
Volt Ch1
Level of Channel 1.
Remote command:
Phase Ch2/1
Phase of Channel 2 relative to Channel 1 (Freq Mode F&Ph).
Remote command:
Freq Ch2
Frequency of Channel 2 (Freq Mode Ch1&2).
Note:
If Gen Track is selected as the measurement time in the analyzer, measurement
channel 2 follows the frequency of generator channel 2. If necessary, the generator fre-
quency is modified slightly so that the signal periods measured during the measure-
ment time contain a sufficient number of samples.
Remote command:
Volt Ch2
Level of Channel 2 (Volt Mode Ch1&2).
Remote command:
Volt Ch2/1
Ratio of the level of both channels relative to Channel 1 (Volt Mode Ch2/1).
Remote command:
Signal change at zero crossing
This function can be selected in all generator instruments, but is available in the analog
generator only with the "Volt Range" = "Fix" setting. Set "Freq Mode" to "Ch1&2".
When there is a change in frequency or level, the signal change occurs
seamlessly
at
the next zero crossing.
Remote command:
Generator Signals
Содержание R&S UPP
Страница 1: ...R S UPP Audio Analyzer User Manual User Manual Version 10 1411105532 ÅP ...
Страница 34: ...Contents R S UPP 18 User Manual 1411 1055 32 10 ...
Страница 630: ...Instrument Functions R S UPP 614 User Manual 1411 1055 32 10 Format ...
Страница 631: ...Instrument Functions R S UPP 615 User Manual 1411 1055 32 10 Format ...