Instrument Functions
R&S
®
UPP
345
User Manual 1411.1055.32 ─ 10
"Time Tick"
Time-controlled measured value recording; measured value triggering
at fixed intervals. In the gap between time ticks, measurements are
started, the measurement results are entered in the measured value
buffer and displayed in the measurement results panel.
This start condition cannot be combined with a time-triggered genera-
tor sweep.
If time tick is combined with the frequency sweep of the selective
RMS measurement or with a synchronous generator sweep, sweep
stepping is performed as usual, but measurement starts only after the
time tick has expired.
If neither the frequency sweep of the selective RMS measurement
nor any generator sweep is active, the time tick defines the sweep x-
axis:
●
During the time tick sweep, the time stamps on the x-axis are
updated and corrected if necessary (e.g. if a measurement lasts
longer than the set time tick) and the individual measured values
are entered according to their correct times.
●
When [START ] is pressed a series of periodic measured value
recordings begins, and when the number of sweep points speci-
fied under "Points" is reached the series starts again from the
beginning.
●
When [SINGLE ] is pressed, a single series of periodic measured
value recordings begins. When the number of sweep points
specified under "Points" is reached the sweep goes to "termina-
ted" status.
Setting Start Condition
Содержание R&S UPP
Страница 1: ...R S UPP Audio Analyzer User Manual User Manual Version 10 1411105532 ÅP ...
Страница 34: ...Contents R S UPP 18 User Manual 1411 1055 32 10 ...
Страница 630: ...Instrument Functions R S UPP 614 User Manual 1411 1055 32 10 Format ...
Страница 631: ...Instrument Functions R S UPP 615 User Manual 1411 1055 32 10 Format ...