Remote Control - Command Description
R&S
®
UPP
699
User Manual 1411.1055.32 ─ 10
SOURce:VOLTage<n2>
<voltage>
Suffix:
<n2>
.
For Multisine generator signals, the suffix selects one of the sin-
ewave 1 to 32 .
For Modulation, suffix 1 or no suffix selects the modulation devi-
ation or depth and suffix 2 the carrier voltage.
For all other generator signals, the suffix is irrelevant and must
be omitted or set to 1.
Parameters:
<voltage>
float
Manual operation:
See
See
See
See
See
See
See
SOURce:VOLTage:EQUalize
<equalize>
Parameters:
<equalize>
ON | OFF
Manual operation:
See
See
See
See
See
See
SOURce:VOLTage:OFFSet<n3>
<offset>
Suffix:
<n3>
.
1|2
The suffix is only relevant for generator function "Stereo Sine". It
selects the channel 1 or 2 for which the offset is entered. The
suffix is irrelevant for all other generator functions and can be
omitted.
Parameters:
<offset>
float
Generator Settings