Instrument Functions
R&S
®
UPP
457
User Manual 1411.1055.32 ─ 10
"RMS"
A broadband RMS measurement is performed which – with the fre-
quency measurement switched on – uses automatic period detection.
(The measurement method in this case corresponds to the "RMS
measurement function " in the "Meas Time Auto mode "). With weak
signals, measurement is performed over a fixed period of 100
ms.
Any superimposed DC components are also measured, but (in the
analog analyzer) are suppressed by the system hardware if the AC
coupling is activated.
Note:
In the case of multitone signals without dominant components,
period detection cannot deliver reliable results; in this case, the RMS
result exhibits slightly larger fluctuations.
"DC"
A DC voltage measurement with automatic AC suppression is per-
formed. With weak signals, measurement is performed over a fixed
period of 100
ms.
In the analog analyzer, DC coupling should be selected for this mea-
surement mode because the DC component of the measurement sig-
nal is filtered out with AC coupling. A suitable warning will be dis-
played if the coupling is switched to AC with DC measurement
enabled, or if the mode is switched to DC measurement with AC cou-
pling selected; however, the measurement is performed.
Note:
In the case of multitone signals without dominant components,
AC period detection cannot deliver reliable results; in this case, the
DC result exhibits slightly larger fluctuations.
"Peak"
A measurement of the input peak value is performed. The numerically
highest input same is determined within a predefined time interval
that is also obtained from the period length of the measurement sig-
nal. With weak signals, measurement is performed over a fixed
period of 100
ms.
Note:
In contrast to peak measurement, which – like all measuring
functions – takes place downstream of the filters, no interpolation fil-
ter (oversampling filter) is used.
Remote command:
OFF | LRMS | DC | PEAK
5.37
Second Monitor
The Second Monitor provides the same selection of functions (RMS, DC or PEAK) as
the Level Monitor or the Input Monitor (see
on page 455). Thus, two different levels (chosen
from RMS, DC or PEAK) in addition to the Function Result (e.g. SINAD) and Fre-
quency can be displayed simultaneously.
2nd Monitor
Switching on Second Monitor and selection of measurement function.
Second Monitor
Содержание R&S UPP
Страница 1: ...R S UPP Audio Analyzer User Manual User Manual Version 10 1411105532 ÅP ...
Страница 34: ...Contents R S UPP 18 User Manual 1411 1055 32 10 ...
Страница 630: ...Instrument Functions R S UPP 614 User Manual 1411 1055 32 10 Format ...
Страница 631: ...Instrument Functions R S UPP 615 User Manual 1411 1055 32 10 Format ...