Instrument Functions
R&S
®
UPP
308
User Manual 1411.1055.32 ─ 10
If the Fix and Lowest modes are selected, the set nominal range value appears in the
next field, where a new range can be set by opening the selection window.
The ranges are graduated in steps of 12 dB from 200 mV to 50 V and allow very good
dynamic range when measuring nonlinear distortion. If "Range Auto" (automatic range
selection) is set, overranging or underranging of the limits in a current measurement
range causes the instrument to switch to the next step. This is also the case for the
"Range Lowest" range setting; here, however, the instrument only goes down to the
selected range limit.
The range values are RMS values for sine waves or other signal shapes up to a crest
factor of
√
2.
CHx = Channel 1 … 8
"Auto"
Automatic measurement range selection. The optimum range is set
separately for each channel.
If an input impedance of 600
Ω
has been selected and the auto
ranger detects a level of more than 12 V in
any channel
, then the
input impedance is switched to 200 k
Ω
for
all channels
and the 50 V
range is activated to protect the input resistors against overload!
"Fix"
The set measurement range is fixed.
If an input impedance of 600
Ω
has been selected,
"Fix" is not avail-
able
for protection of the input resistors against overload!
Note:
If a voltage exceeding the selected level range is applied, the
analyzer status is displayed with a red digit in the lower status line
(the digit indicates the channel that is overranged) and the measured
values are shown as invalid ("---"). On switching from Auto to Fix, the
currently valid range in the channel is taken over.
"Lowest"
The set measurement range is fixed as the lowest range; in the event
of overdrive, this range is automatically substituted by higher ranges.
If an input impedance of 600
Ω
was selected,
the 50 V measurement
range is not available
. If the auto ranger detects a level of more
than 12 V in any channel, then the input impedance is switched to
200
k
Ω
for all channels and the 50 V range is activated to protect the
input resistors against overload.
Note:
On switching from Auto to Lowest, the currently valid range in
the channel is taken over.
Remote command:
SENSe[1]:VOLTage:RANGe<n3>:MODE
SENSe[1]:VOLTage:RANGe<n3>:VALue
Ref Imped
Reference value for impedance.
Configure Analog Analyzer
Содержание R&S UPP
Страница 1: ...R S UPP Audio Analyzer User Manual User Manual Version 10 1411105532 ÅP ...
Страница 34: ...Contents R S UPP 18 User Manual 1411 1055 32 10 ...
Страница 630: ...Instrument Functions R S UPP 614 User Manual 1411 1055 32 10 Format ...
Страница 631: ...Instrument Functions R S UPP 615 User Manual 1411 1055 32 10 Format ...