Instrument Functions
R&S
®
UPP
314
User Manual 1411.1055.32 ─ 10
After the Start Cond has been selected, measurement either begins immediately (Auto)
or only when the desired trigger condition is fulfilled; for details see
Delay
Delay for Start Cond.
A delay time can be entered for the Start Cond Auto menu item; for details see
Max FFT Size
Used to limit automatic zoom fitting for the FFT size for all FFT-supported measure-
ment functions (e.g. THD, THD+N, DFD, each with Post FFT) to the selected value
with the goal of increasing measurement speed.
The limit to the FFT size is particularly useful for external sweeps for adapting the mea-
suring time to the time frame of an external time tick. A limit is usually associated with
degradation in measuring accuracy since incidents such as carrier-level noise can no
longer be completely recorded.
Remote command:
5.12.2
Interface Assignment
Use of the R&S
UPP-B2 (digital audio interfaces) option for 2-channel analysis.
DIGITAL BAL plug (9-pin D-Sub connector)
Configuring the Digital Analyzer