Remote Control - Command Description
R&S
®
UPP
743
User Manual 1411.1055.32 ─ 10
Example:
First example:
TRAC:SWEep1:LOAD:AY?
Outputs the data of the Y-axis of channel A from the first sweep
system.
Response with the setting
FORMat ASC
:
-57.1647,-6.08881,-32.304,-53.6429 …
Response with the setting
FORMat REAL
:
#<Lengthoflength><length><block_data>
Second example
TRACe:SWEep1:LOAD:COUNt:BY?
Outputs the number of data of the Y-axis of channel B from the
first sweep subsystem
Response:
50
Usage:
Query only
TRACe:FFT<n2>:LOAD:AX?
TRACe:FFT<n2>:LOAD:AY?
TRACe:FFT<n2>:LOAD:BX?
TRACe:FFT<n2>:LOAD:BY?
TRACe:FFT<n2>:LOAD:COUNt:AX?
TRACe:FFT<n2>:LOAD:COUNt:AY?
TRACe:FFT<n2>:LOAD:COUNt:BX?
TRACe:FFT<n2>:LOAD:COUNt:BY?
Commands for outputting FFT data and outputting the number of FFT data. See also
TRACe:SWEep<n2>:LOAD:COUNt:BY?
Suffix:
<n2>
.
1 | 2
subsystems 1 and 2
Return values:
<argument>
integer | float | binary
For the
COUNt
commands, an integer value is returned.
The format of the return values of the other commands depends
on the
FORMat
setting:
If
FORMat ASCii
is selected float values are returned.
If
FORMat REAL
is selected a binary data block is returned.
Usage:
Query only
TRACe:WAVeform<n2>:LOAD:AX?
TRACe:WAVeform<n2>:LOAD:AY?
TRACe:WAVeform<n2>:LOAD:BX?
TRACe:WAVeform<n2>:LOAD:BY?
TRACe:WAVeform<n2>:LOAD:COUNt:AX?
TRACe:WAVeform<n2>:LOAD:COUNt:AY?
Triggering Measurements and Reading Results
Содержание R&S UPP
Страница 1: ...R S UPP Audio Analyzer User Manual User Manual Version 10 1411105532 ÅP ...
Страница 34: ...Contents R S UPP 18 User Manual 1411 1055 32 10 ...
Страница 630: ...Instrument Functions R S UPP 614 User Manual 1411 1055 32 10 Format ...
Страница 631: ...Instrument Functions R S UPP 615 User Manual 1411 1055 32 10 Format ...