Manual Operation
R&S
®
UPP
142
User Manual 1411.1055.32 ─ 10
Below is an example for entering the generator output voltage: Additional units can be
selected with the >> button:
4.7.14
Completing Settings
For the entry of numeric values, the R&S
UPP distinguishes between Edit mode (iden-
tified by a green selection frame) and Direct mode (identified by a magenta selection
frame). See also the section
Chapter 4.7.9, "Numeric Entries – Numeric Entry Fields"
●
Numeric entries in
Edit mode
are adopted and set in the hardware only by using
[ENTER] or selecting a unit via the softkey bar.
●
In
Direct mode
, numeric values entered or changed using the mouse wheel are
immediately passed on to the hardware and executed.
With checkboxes, functions are activated and deactivated by the hardware as soon as
the status of the checkbox is toggled with the ENTER function. For operation using a
mouse, functions are activated/deactivated as soon as the checkbox is clicked.
With radio buttons, settings are executed by the hardware as soon as the radio button
is toggled with the ENTER function. For operation using a mouse, functions are tog-
gled as soon as the corresponding radio button is clicked.
A parameter from a selection field is also executed by activating the ENTER function.
In this case, a simple mouse click will select and activate the desired parameter.
Most settings are made without any noticeable adjustment or calculation times. If a
brief calculation time is necessary (e.g. to calculate a multi-tone signal adapted to the
FFT analysis), the calculation is performed in the background while ongoing measure-
ments continue.
Function
Computer keyboard
Mouse
Finishing numeric value entry
(Edit mode)
Complete entry with the ENTER
key or
Use the F5 to F12 keys to select
the desired unit or
Complete the entry with the m, k,
etc. keys to enter the value, e.g.
as mV or kHz.
Click the softkey button with the
desired unit.
Immediately setting numeric value
changes in the hardware
(Direct mode)
Highlight the numeric entry field,
switch to Direct mode with the
Enter key, position the digit cursor
with the > or < cursor keys and
enter the new digit directly using
the numeric keys.
---
Settings in the Panels
Содержание R&S UPP
Страница 1: ...R S UPP Audio Analyzer User Manual User Manual Version 10 1411105532 ÅP ...
Страница 34: ...Contents R S UPP 18 User Manual 1411 1055 32 10 ...
Страница 630: ...Instrument Functions R S UPP 614 User Manual 1411 1055 32 10 Format ...
Страница 631: ...Instrument Functions R S UPP 615 User Manual 1411 1055 32 10 Format ...