Instrument Functions
R&S
®
UPP
606
User Manual 1411.1055.32 ─ 10
If the file contains both reference traces and limit traces,
File Intern
is displayed in the
corresponding parameter fields. "File Intern" means that there is no individual value
(for example in the case of a "Reference", this would be "Limit Upper" or "Limit Lower"
with the setting Value) but instead a trace, and that this trace has been taken from a
file loaded at Y-Source = File:
Trace
A or File:
Trace
B.
Furthermore, all Auto and Tracking checkboxes such as Label Auto, Unit Func Track,
Unit Auto etc. are deactivated and grayed out so that they cannot be used. All parame-
ter fields that allow Auto settings, such as "X-Source " or "Scaling ", are set to "Man-
ual", since they are preset by the file content.
Load