Maintenance and Instrument Interfaces
R&S
®
UPP
879
User Manual 1411.1055.32 ─ 10
Damage of instrument when using wrong fuses
Using wrong fuses may cause damage of the instrument.
Changing the AC supply fuses:
1. Disconnect the power cable.
2. Pry out the fuse holder with two small screwdrivers or similar tools.
3. Remove both fuses and install new ones.
4. Reinsert the fuse holder.
9.4
Instrument Interfaces
The following section describes the remote-control interfaces of the audio analyzer and
the pin assignment of the monitor socket. The other interfaces are described in
9.4.1
IEC/IEEE-Bus Interface (IEC
625 / IEEE
488)
As standard, the instrument is equipped with an IEC/IEEE-bus connector. The connec-
tor socket (conforming to IEEE
488) is at the rear of the instrument. The interface can
be used to connect a controller to allow remote control of the instrument. A shielded
cable is used for connection.
9.4.1.1
Features of the IEC/IEEE-bus interface
●
8-bit parallel data transfer
●
Bidirectional data transfer
●
Three-wire handshake
●
High data transfer rate
●
Up to 15 instruments can be connected
●
Max. length of connecting cables: 15
m (individual connection 2
m)
●
Wired-Or operation with parallel connection of several instruments