Instrument Functions
R&S
®
UPP
357
User Manual 1411.1055.32 ─ 10
Termination event
… stops the current sweep immediately and causes the single sweep to be restarted
(see
Chapter 5.20.1, "Overview of Sweep Options"
X stop value reached
The single sweep is stopped; the sweep system changes to the "Sweep Waiting" state.
Remote command:
-
5.20.3.3
Terminated sweep
The sweep system only enters this state if a single sweep or a 2-dimensional continu-
ous sweep is terminated. No further measurements are performed; the measured val-
ues, data sets and sweep traces are frozen. Sweep traces should only be read out via
the remote-control station when the system is in this state.
The sweep state is indicated by "Sweep Waiting ".
A terminated single sweep is affected by the following events:
START
Pressing [START ]
●
deletes the sweep curves (in multiscan mode all scans are deleted) including the
min/max curves
●
resets the Min/Max. data and the limit violations of the sweep curves
●
starts a continuous sweep
Remote command:
ON
STARt
SINGLE
Pressing [SINGLE ] causes the single sweep to be restarted.
Remote command:
ON
SINGle
STOP/CONT
Pressing [STOP/CONT ] starts continuous measurement. Reactivating the single
sweep with [SINGLE ].
Remote command:
-
Termination event
… starts the continuous measurement (see
Chapter 5.20.1, "Overview of Sweep
Sweep Options
Содержание R&S UPP
Страница 1: ...R S UPP Audio Analyzer User Manual User Manual Version 10 1411105532 ÅP ...
Страница 34: ...Contents R S UPP 18 User Manual 1411 1055 32 10 ...
Страница 630: ...Instrument Functions R S UPP 614 User Manual 1411 1055 32 10 Format ...
Страница 631: ...Instrument Functions R S UPP 615 User Manual 1411 1055 32 10 Format ...