Getting Started
R&S
®
UPP
97
User Manual 1411.1055.32 ─ 10
The panels for configuring the measured value display are used to determine how
the measurement results are presented (i.e. which unit, how many digits, etc.). A
series of panels for the different measurement windows are provided for this pur-
pose.
The "Function
Config" panel is shown here as an example; the first column of the
numeric measurement display is configured in this panel, i.e. the display fields
which show the measured level values for the frequency response measurement
described here.
The unit for the numeric measurement displays is set in the Unit field.
The Engineering parameter is used to determine whether the display automatically
toggles between µV, mV and V, for example, or whether other unit formats are
desired.
If the measurement results are to be displayed as relative values, they are referred
to the reference value set in the Reference field.
Other parameter fields in this panel are used to set limits and to display extreme
values. These functions are described in detail in
Chapter 5, "Instrument Functions"
5. Step: Starting the sweep
Although the analyzer is already supplying measurement results, the status mes-
sage "Sweep Waiting" above the softkeys indicates that the sweep has not yet
been started although it has been configured in the "Generator Function" panel.
a) This is done with the "Start" button from the toolbar (the two double arrows) or
the "Ctrl F5" key on the external keyboard. The status message "Sweep Run
Cont" informs the user that when this key is pressed the sweep will start and be
repeated continuously.
b) The "Stop/Cont" button of the toolbar (two vertical bars) or the "Ctrl F7" key of
the external keyboard can be used to interrupt the sweep process at any time
("Sweep Stopped" status message), pressing the key again continues the proc-
ess.
Introduction to Instrument Operation Using Examples
Содержание R&S UPP
Страница 1: ...R S UPP Audio Analyzer User Manual User Manual Version 10 1411105532 ÅP ...
Страница 34: ...Contents R S UPP 18 User Manual 1411 1055 32 10 ...
Страница 630: ...Instrument Functions R S UPP 614 User Manual 1411 1055 32 10 Format ...
Страница 631: ...Instrument Functions R S UPP 615 User Manual 1411 1055 32 10 Format ...