Instrument Functions
R&S
®
UPP
283
User Manual 1411.1055.32 ─ 10
5.8.2
Setting Bytes as Numerical Values
A byte can be set to any value in this entry block. It is even possible to set bit combina-
tions that are not yet defined in the standard.
Byte Number
Selects the byte, 0 to 4.
Remote command:
SOURce:PROTocol:NUMerical:BYTE
SOURce:PROTocol:CH<n3>:BYTE<n4>
Value
The decimal value for the byte selected is displayed here and can also be entered.
Remote command:
SOURce:PROTocol:NUMerical:VALue
SOURce:PROTocol:CH<n3>:BYTE<n4>
Hex Entry
The byte value can be displayed or entered in hexadecimal, binary or decimal.
If the entry brings about a bit combination that is not yet defined in the standard, the
parameter has no meaning that can be displayed. In this event *********** is displayed.
5.8.3
Setting Bytes in Remote-Control Mode
When entering or querying values in remote control mode, an easier method than the
two individual commands for byte and value
SOURce:PROTocol:NUMerical:BYTE <y>
SOURce:PROTocol:NUMerical:VALue <n>
is to use the following, combined remote-control command:
SOURce:PROTocol:CH<x>:BYTE<y> <n>
for the input
SOURce:PROTocol:CH<x>:BYTE<y>?
for a query
After the switchover back into Local mode, the settings which are configured via this
remote-control command in the protocol generator panel are displayed in plain text in
the corresponding parameter fields.
"Parameters:"
<x>
→
1
<y>
→
0 to 3
Protocol Generator
Содержание R&S UPP
Страница 1: ...R S UPP Audio Analyzer User Manual User Manual Version 10 1411105532 ÅP ...
Страница 34: ...Contents R S UPP 18 User Manual 1411 1055 32 10 ...
Страница 630: ...Instrument Functions R S UPP 614 User Manual 1411 1055 32 10 Format ...
Страница 631: ...Instrument Functions R S UPP 615 User Manual 1411 1055 32 10 Format ...