Getting Started
R&S
®
UPP
95
User Manual 1411.1055.32 ─ 10
a) The basic settings for configuring the analyzer are made in the "Analyzer
Con-
fig" panel.
In order to operate the panel, it must be selected and have the focus.
The frequency response is usually measured on all channels; this is already
configured in the Meas Channel field. The following parameter fields occur only
once and are set to "Track" and affect all channels.
b) The "AC/DC Coupl" parameter field is set to "Track", i.e. AC and DC coupling
can be switched as necessary for all channels using the two radio buttons. If
the coupling is to be set individually for each channel, then the AC/DC Coupl
parameter field must be set to Split.
The fields for the measurement bandwidth are already preset with the appropri-
ate setting; no prefilter is required here.
The signal to be measured is fed via the balanced input sockets.
The auto-range function for all channels ("Range = Track") is generally the cor-
rect setting here. If range is to be set individually for each channel, then the
"Range" parameter field must be set to "Split".
The default settings in the following fields are also correct.
This completes the settings in the "Analyzer
Config" panel.
Introduction to Instrument Operation Using Examples