Instrument Functions
R&S
®
UPP
254
User Manual 1411.1055.32 ─ 10
Volt RMS
Used to enter the RMS value of the noise signal (only in the analog generator). The
peak value to RMS value ratio of the signal (crest factor) is minimized. If the noise is
defined in the time domain, the selected distribution function for amplitude determines
the crest factor.
Remote command:
Filter
The generator signal is filtered in the time domain. One of the user
‑
definable filters in
the Filter panel or a weighting filter can be selected. All level settings apply to the unfil-
tered signal.
Remote command:
For a description of the filters and their setting parameters, see
Equalizer
The generator signal is equalized in the frequency domain using the file selected in the
"Equal File" parameter field. The checkbox enables or disables this function.
These parameter fields are hidden if time-domain noise (Domain Time) is selected.
The button opens a browser in which you can select the equalizer file, refer also to
Remote command:
DC Offset
A DC offset is superimposed on the generator signal. The DC offset decreases the
maximum AC voltage. The checkbox enables or disables this function. The value con-
figured for the offset voltage is always displayed.
Remote command:
SOURce:VOLTage:OFFSet<n3>:STATe
Generator Signals
Содержание R&S UPP
Страница 1: ...R S UPP Audio Analyzer User Manual User Manual Version 10 1411105532 ÅP ...
Страница 34: ...Contents R S UPP 18 User Manual 1411 1055 32 10 ...
Страница 630: ...Instrument Functions R S UPP 614 User Manual 1411 1055 32 10 Format ...
Страница 631: ...Instrument Functions R S UPP 615 User Manual 1411 1055 32 10 Format ...