Instrument Functions
R&S
®
UPP
551
User Manual 1411.1055.32 ─ 10
Sweep, FFT, Waveform
"Manual"
Manual scaling
The scaling of a graphical display is determined using the "Left " and
"Right " parameter fields below.
Each time the "Manual " parameter is set, the graphical display is
deleted and then replotted with the scale end values entered under
"Left " and "Right ".
"Auto"
Automatic scaling
Auto Scale uses suitable default values for
one-off
scaling of the x-
axis. These values depend on the measurement task and the subsys-
tem (e.g. the start/stop values of the sweep or FFT are used in the
generator and analyzer panel).
The entry fields for the "Left " and "Right " parameter fields below are
displayed in gray (cannot be accessed) with the obtained values.
Remote command:
Spacing
Linear or logarithmic spacing of the x-axis.
For logarithmic units (dB units) and linear units in which negative numbers can occur
(e.g. ∆V, ∆W), this parameter field is displayed as not user-accessible and is set to
"Lin", since it makes no sense / is not possible to display dB values or negative values
logarithmically.
Available in the graphics subsystems
Sweep, FFT, Waveform
"Lin"
Linear spacing of the x-axis.
"Log"
Logarithmic spacing of the x-axis.
Remote command:
Left
Used to set the left-hand value on the x-axis.
This parameter field can only be accessed if "Scaling"
=
"Manual" is selected.
Negative values or the value 0.00000 are not permitted if "Spacing"
=
"Log".
Available in the graphics subsystems
Sweep, FFT, Waveform
Remote command:
Right
Curve, Spectrum and Bar Graph Display
Содержание R&S UPP
Страница 1: ...R S UPP Audio Analyzer User Manual User Manual Version 10 1411105532 ÅP ...
Страница 34: ...Contents R S UPP 18 User Manual 1411 1055 32 10 ...
Страница 630: ...Instrument Functions R S UPP 614 User Manual 1411 1055 32 10 Format ...
Страница 631: ...Instrument Functions R S UPP 615 User Manual 1411 1055 32 10 Format ...