Remote Control - Command Description
R&S
®
UPP
750
User Manual 1411.1055.32 ─ 10
7.5.7
Output of Protocol Data
SENSe<8>:PROTocol:CH<1|2>:BYTE<0...4>?
Outputs the protocol data of the selected byte of the selected channel.
Suffix:
<8>
.
8 = protocol analysis
<1|2>
selects channel 1 or 2
<0...4>
selects byte 0 to 4
Return values:
<argument>
0 to 255
Return value = decimal bit equivalent
Example:
SENS8:PROT:CH1:BYTE2?
Outputs the protocol data of byte 2 (the third byte) of channel 1.
Usage:
Query only
7.5.8
Output of Protocol Errors
●
●
SENSe<8>:PROTocol:ERRor:CRC<1|2>?
●
SENSe<8>:PROTocol:ERRor:INV<1|2>?
●
SENSe<8>:PROTocol:ERRor:LOC<1|2>?
●
SENSe<8>:PROTocol:ERRor:PAR<1|2>?
●
SENSe<8>:PROTocol:ERRor:PCM<1|2>?
SENSe<8>:PROTocol:ERRor?
Suffix:
<8>
.
8 = protocol analysis
Return values:
<argument>
integer
return values:
0,"No error"
or
<n>,"PCM1,PCM2,PAR1,PAR2,…"
The entire string is not returned in quotation marks, but only the
mnemonic part.
<n> represents 10 bits (d0 to d9)
<n> = 0 to 1023
The individual errors are represented by means of bits in the
integer value <n>:
d0: PCM1
d1: PCM2
d2: PAR1
Triggering Measurements and Reading Results