220
────────────────────────────────────────────────────
12.8 Notes on Interface
────────────────────────────────────────────────────
12.8.2 RS-232C Troubleshooting
Symptom
Cause / Treatment
The RS-232C has stopped
working completely.
Are the cables properly connected?
Are all the devices powered on?
Are the cables properly connected?
Is the STATUS screen or the FDD screen on?
Transmission on the RS-232C is
not taking place properly.
Is the controller message terminator set correctly?
(TRAN:TERM command)
(Refer to "Message Terminators")
Is RS-232C (band rate, data length, parity, stop bits) set the
same?
After transmission on the RS-
232C, the keys on this unit freeze
up and have no effect.
Press the LOCAL key on the front panel of this unit to release the
remote state.
When attempting to read data
using a Basic INPUT statement,
the RS-232C hangs.
Be sure to transmit one query before each Basic INPUT
statement.
Have any of these transmitted queries resulted in an error?
Although a command has been
transmitted, nothing has
happened.
Using the *ESR? query, inspect the standard event status register,
and check what type of error has occurred.
Sending several queries, produces
only one response.
Has an error occurred?
Send the queries one at a time, and read the responses
individually.
When you want to read them in all at once, try doing so by
putting them all on one line separated by the message separator
character.
Have you used the *IDN? query?
The response message to a query
differs from the display on the
front panel of this unit.
Due to the response message being produced at the instant that
this unit receives the query, there is a possibility that it may not
agree with the display at the instant that the controller reads it in.
Summary of Contents for Power HiTester 3193
Page 2: ......
Page 50: ...32 3 9 Operations During Power Failure ...
Page 76: ...58 4 13 Degaussing ...
Page 80: ...62 5 2 Setting the Frequency Range fa ...
Page 108: ...90 9 3 Internal Circuit for the External Control and Timing ...
Page 112: ...94 10 3 Output Rate ...
Page 250: ...232 13 9 Error and Overflow Displays ...
Page 278: ...260 17 5 Internal Block Diagram ...
Page 284: ...266 19 2 Installation Procedures For JIS standard For EIA standard External Dimensions ...
Page 300: ...282 20 4 Internal Block Diagram of the 3193 ...
Page 306: ...INDEX 4 Index ...
Page 307: ......
Page 308: ......
Page 309: ......
Page 310: ......