7.8 Zero suppress function
────────────────────────────────────────────────────
F1
F2
INTEG-ZERO
F3
NOTE
7.8 Zero suppress function
You can set the minimum value of data to be integrated. You can change
the value if the input level is too low for the preset range.
1. Press the
STATUS
key, then use the
PAGE
(
)
keys to display the "
SYSTEM
" page.
2. Using the
CURSOR
keys, move the cursor to the
"
INTEG-ZERO
" item.
3. Select from
F1
(0.5%)
,
F2
(0.1%)
,
F3
(0.0%)
.
4. Press the
MEAS
key to return to the measurement
screen.
・
In zero suppress, figures less than the percentage set for the full range are
regarded as "0".
・
The setting is 0.5% by default.
・
When the setting is at 0.0% or 0.1%, the values not displayed as
instantaneous values will also be counted. Counts may be displayed even
when there is no input due to internal off-set.
・
This change function is not transferred when data is saved or loaded via PC
communications or floppy disk. Therefore, settings must be made manually
before measurement.
・
The zero suppress in the display of instantaneous values is not changed.
Summary of Contents for Power HiTester 3193
Page 2: ......
Page 50: ...32 3 9 Operations During Power Failure ...
Page 76: ...58 4 13 Degaussing ...
Page 80: ...62 5 2 Setting the Frequency Range fa ...
Page 108: ...90 9 3 Internal Circuit for the External Control and Timing ...
Page 112: ...94 10 3 Output Rate ...
Page 250: ...232 13 9 Error and Overflow Displays ...
Page 278: ...260 17 5 Internal Block Diagram ...
Page 284: ...266 19 2 Installation Procedures For JIS standard For EIA standard External Dimensions ...
Page 300: ...282 20 4 Internal Block Diagram of the 3193 ...
Page 306: ...INDEX 4 Index ...
Page 307: ......
Page 308: ......
Page 309: ......
Page 310: ......