149
────────────────────────────────────────────────────
:CURRent [channel no.] :RANGe?
────────────────────────────────────────────────────
Syntax
:CURRent[
1 - 6
]:RANGe?
Response
syntax
:CURRENT[
1 - 6
]:RANGE
<NR1>
Example
Transmission
Response
:CURRENT1:RANGE?
:CURRENT1:RANGE 50
Syntax
:CURRent
[1 - 6]
?
Response
syntax
:CURRENT
[1 - 6]
:AUTO <
ON/OFF
>;
MEAN <
ON/OFF
>;RANGE
<NR1>
Example
Transmission
Response
:CURRENT1?
:CURRENT1:AUTO ON;MEAN ON;RANGE
10
Syntax
:DATAout?
Response
syntax
:DATAOUT:ITEM:
NORMAL
<0-63 (8 items)>
;
SUM
<0-63 (7 items)>
;
INTEGRATE
<0-63 (10 items)>
;
FREQUENCY
<0-7>
;
LOADFACTOR
<0-64 (2 items)>
;
EFFICIENCY
<0-7>
;
EXTERNALIN
<0-7>
;
FD
<ON/OFF>
;
PRINTER
<ON/OFF>
Example
Transmission
Response
:DATAOUT:ITEM?
:DATAOUT:ITEM:NORMAL 7,7,0,0,0,
0,0,0;SUM 8,8,8,8,8,8,8,0;
INTEGRATE 0,0,7,0,0,0,8,8,8,1;
FREQUENCY 1;LOADFACTOR 0,8;
EFFICIENCY 1;EXTERNALIN 7;
FD ON;PRINTER OFF
Function
Queries the current ranging of the
specified input unit.
Function
Queries the current settings for the
specified input unit.
Function
Queries the all setting items on a FDD or
printer.
:CURRent [
channel no.
] :RANGe?
Queries the current ranging.
:CURRent [
channel no.
]?
Queries the current measurement.
:DATAout?
Queries the all setting items on a FDD or printer.
Summary of Contents for Power HiTester 3193
Page 2: ......
Page 50: ...32 3 9 Operations During Power Failure ...
Page 76: ...58 4 13 Degaussing ...
Page 80: ...62 5 2 Setting the Frequency Range fa ...
Page 108: ...90 9 3 Internal Circuit for the External Control and Timing ...
Page 112: ...94 10 3 Output Rate ...
Page 250: ...232 13 9 Error and Overflow Displays ...
Page 278: ...260 17 5 Internal Block Diagram ...
Page 284: ...266 19 2 Installation Procedures For JIS standard For EIA standard External Dimensions ...
Page 300: ...282 20 4 Internal Block Diagram of the 3193 ...
Page 306: ...INDEX 4 Index ...
Page 307: ......
Page 308: ......
Page 309: ......
Page 310: ......