:MEASure:ITEM:LOADfactor
────────────────────────────────────────────────────
Syntax
:MEASure:ITEM:LOADfactor
<NR1, (2 items)>
<NR1> = 0 - 63
Example
:MEASURE:ITEM:LOADFACTOR 7,0
Sets the measurement items of the load
factor to LF1, LF2, LF3.
Syntax
:MEASure:ITEM:LOADfactor?
Response
syntax
:MEASURE:ITEM:LOADFACTOR
<NR1,(2items)>
<NR1>
= 0 ‑ 63
Example
Transmission
Response
:MEASURE:ITEM:LOADFACTOR?
:MEASURE:ITEM:LOADFACTOR 7,0
Function
Sets the default items (only load factor
LF
) to be transferred in the response
message to the :MEASure? query in the
default mode.
The items are set as shown below by
setting bits, to specify two numerical
values.
bit
7
bit
6
bit
5
bit
4
bit
3
bit
2
bit
1
bit
0
1
- -
LF6
LF5
LF4
LF3
LF2
LF1
2
- -
LF45
LF456
LF123
LF56
LF34
LF12
Error
Execution error / If the setting data is out
of the range.
Note
Depending on connection mode and
optional unit, the item which is not
displayed can not be selected. The set
items which cannot be selected is
ignored.
Function
Queries the output items specified by the
:MEASure:ITEM:LOADfactor <NR1>
:MEASure:ITEM:LOADfactor
Sets the output item for the load factor.
:MEASure:ITEM:LOADfactor?
Queries the output item for the load factor?
Summary of Contents for Power HiTester 3193
Page 2: ......
Page 50: ...32 3 9 Operations During Power Failure ...
Page 76: ...58 4 13 Degaussing ...
Page 80: ...62 5 2 Setting the Frequency Range fa ...
Page 108: ...90 9 3 Internal Circuit for the External Control and Timing ...
Page 112: ...94 10 3 Output Rate ...
Page 250: ...232 13 9 Error and Overflow Displays ...
Page 278: ...260 17 5 Internal Block Diagram ...
Page 284: ...266 19 2 Installation Procedures For JIS standard For EIA standard External Dimensions ...
Page 300: ...282 20 4 Internal Block Diagram of the 3193 ...
Page 306: ...INDEX 4 Index ...
Page 307: ......
Page 308: ......
Page 309: ......
Page 310: ......