APPENDIX 2
────────────────────────────────────────────────────
────────────────────────────────────────────────────
Voltage
Load
②
①
Diagram for the example
using the option 9605
3V3A wiring
①
Based on 3P3W wiring, the three channels of voltage and current are
summed. The figures above show the state when the load is purely resistive.
②
With an inductive load such as a motor
The effective power of ch1 appears negative because I1 lags U1 by more
than 90
°
. However, the summed data for ch(3) is not related to power
measurement, so wiring can be reversed.