165
────────────────────────────────────────────────────
:FREQuency [channel no.] :SOURce?
────────────────────────────────────────────────────
Syntax
:FREQuency
[A/B/C]
:SOURce
<A>
<A> = U1, U2, U3, U4, U5, U6, I1,
I2, I3, I4 ,I5, I6
If the setting of external input
for channel B is pulse
measurement, the frequency
measurement source for
channel C is forcibly set to
"EXTB" and changing is not
possible.
Example
:FREQUENCYA:SOURCE U1
Sets the frequency measurement source
for channel A to U1.
Syntax
:FREQuency
[A/B/C]
:SOURce?
Example
Transmission
Response
:FREQUENCYA:SOURCE?
:FREQUENCYA:SOURCE U1
Syntax
:FREQuency
[A/B/C]
?
Response
syntax
:FREQUENCY
[A/ B/ C]
:AUTO
<ON/OFF>
;RANGE
<1/5/10>
;SOURCE
<A>
Example
Transmission
Response
:FREQUENCYA?
:FREQUENCYA:AUTO ON;RANGE 50;
SOURCE U1
Function
Sets the source of the frequency
measurement for specified channel.
Function
Queries the current setting of the
frequency source for the specified
channel.
Function
Queries the settings for the specified
channel of the frequency measurement.
:FREQuency [
channel no.
] :SOURce
Sets the channel source of the frequency measurement.
:FREQuency [
channel no.
] :SOURce?
Queries the channel source of the frequency measurement.
:FREQuency [
channel no.
] ?
Queries the settings for frequency measurement channel.
Summary of Contents for Power HiTester 3193
Page 2: ......
Page 50: ...32 3 9 Operations During Power Failure ...
Page 76: ...58 4 13 Degaussing ...
Page 80: ...62 5 2 Setting the Frequency Range fa ...
Page 108: ...90 9 3 Internal Circuit for the External Control and Timing ...
Page 112: ...94 10 3 Output Rate ...
Page 250: ...232 13 9 Error and Overflow Displays ...
Page 278: ...260 17 5 Internal Block Diagram ...
Page 284: ...266 19 2 Installation Procedures For JIS standard For EIA standard External Dimensions ...
Page 300: ...282 20 4 Internal Block Diagram of the 3193 ...
Page 306: ...INDEX 4 Index ...
Page 307: ......
Page 308: ......
Page 309: ......
Page 310: ......