177
────────────────────────────────────────────────────
:MEASure:ITEM:SUM
────────────────────────────────────────────────────
Syntax
:MEASure:ITEM:SUM
<NR1, (7
items)>
<NR1> = 0 - 63.
Example
:MEASURE:ITEM:SUM 7,7,7,0,0, 0,0
Sets the U12, U34, U56, I12, I34, I56,
P112, P34, P56 as the default items.
Syntax
:MEASure:ITEM:SUM?
Response
syntax
:MEASURE:ITEM:SUM
<0-63>
Example
Transmission
Response
:MEASURE:ITEM:SUM?
:MEASURE:ITEM:SUM 7,7,7,0,0,0,0
Function
Sets the default items (only SUM value)
to be transferred in the response message
to the :MEASure? query in the default
mode.
The items are set as shown below by
setting bits, to specify seven numerical
values.
bit
7
bit
6
bit
5
bit
4
bit
3
bit
2
bit
1
bit
0
1 - -
U45
U456
U123
U56
U34
U12
2 - -
I45
I456
I123
I56
I34
I12
3 - -
P45
P456
P123
P56
P34
P12
4 - -
S45
S456
S123
S56
S34
S12
5 - - Q45
Q456
Q123
Q56
Q34
Q12
6 - - PF45 PF456 PF123 PF56
PF34
PF12
7 - - DEG45 DEG456 DEG123 DEG56 DEG34 DEG12
Error
Execution error / If the setting data is out
of the range.
Note
Depending on the optional unit installing,
if the set items which cannot be selected
is ignored.
Function
Queries the default items specified by
SUM value.
:MEASure:ITEM:SUM
Sets the output item for the SUM value.
:MEASure:ITEM:SUM?
Queries the output item for the SUM value.
Summary of Contents for Power HiTester 3193
Page 2: ......
Page 50: ...32 3 9 Operations During Power Failure ...
Page 76: ...58 4 13 Degaussing ...
Page 80: ...62 5 2 Setting the Frequency Range fa ...
Page 108: ...90 9 3 Internal Circuit for the External Control and Timing ...
Page 112: ...94 10 3 Output Rate ...
Page 250: ...232 13 9 Error and Overflow Displays ...
Page 278: ...260 17 5 Internal Block Diagram ...
Page 284: ...266 19 2 Installation Procedures For JIS standard For EIA standard External Dimensions ...
Page 300: ...282 20 4 Internal Block Diagram of the 3193 ...
Page 306: ...INDEX 4 Index ...
Page 307: ......
Page 308: ......
Page 309: ......
Page 310: ......