84
────────────────────────────────────────────────────
8.4 Example Measurement
────────────────────────────────────────────────────
8.4.5 Efficiency Measurement of an Inverter (3
φ
3W) and Motor
Input side (3φ3W)
50/60 Hz
P
123
P
456
Output side (3φ3W)
Basic wave: 0.5-100 Hz
Carrier: 2-20 kHz
Inverter
Torque
meter
Motor
Number
of rotating
(chB)
Torque value
(chA)
NOTE
When channel 1, 2, 3 are the input and channel 4, 5, 6 the output:
Channel 1, 2, 3: input side, channel 4, 5, 6: output side
Input: 3V3A, AC mode / output: 3V3A, AC or AC+DC mode
Analog output from the torque meter: to channel A of the 9603
Analog output from the rotation counter: to channel B of the 9603
Motor power:
P
m
・
Efficiency of an inverter :
P
456
η
1 = ─── ×100
P
123
・
Efficiency of a motor :
P
m
η
2 = ─── ×100
P
456
・
Efficiency of total :
P
m
η
3 = ─── ×100
P
123
Measuring the efficiency of an inverter (3
φ
3W) and motor
The torque meter and rotation counter analog outputs used should be chosen
to have as fast as possible a response.
Summary of Contents for Power HiTester 3193
Page 2: ......
Page 50: ...32 3 9 Operations During Power Failure ...
Page 76: ...58 4 13 Degaussing ...
Page 80: ...62 5 2 Setting the Frequency Range fa ...
Page 108: ...90 9 3 Internal Circuit for the External Control and Timing ...
Page 112: ...94 10 3 Output Rate ...
Page 250: ...232 13 9 Error and Overflow Displays ...
Page 278: ...260 17 5 Internal Block Diagram ...
Page 284: ...266 19 2 Installation Procedures For JIS standard For EIA standard External Dimensions ...
Page 300: ...282 20 4 Internal Block Diagram of the 3193 ...
Page 306: ...INDEX 4 Index ...
Page 307: ......
Page 308: ......
Page 309: ......
Page 310: ......