148
────────────────────────────────────────────────────
:CURRent [channel no.] :MEAN
────────────────────────────────────────────────────
Syntax
:CURRent
[1 - 6]
:MEAN
<ON/OFF>
Example
:CURRENT1:MEAN ON
Sets the rectifier type of current
ranging for channel 1 of the input unit
to MEAN.
Syntax
:CURRent
[1 - 6]
:MEAN?
Response
syntax
:CURRENT
[1 - 6]
:MEAN
<ON/OFF>
Example
Transmission
Response
:CURRENT1:MEAN?
:CURRENT1:MEAN ON
Syntax
:CURRent
[1 - 6]
:RANGe
<NR1>
<NR1> =
When using the 9600, 9601:
0.2, 0.5, 1, 2, 5, 10, 20, 50
When using the 9602 and 20 A sensor:
0.5, 1, 2, 5, 10, 20
When using the 9602 and CT6862:
1.25, 2.5, 5, 12.5, 25, 50
(CT ratio=2.5)
0.5, 1, 2, 5, 10, 20
(CT ratio=other than 2.5)
When using the 9602 and 200 A sensor:
5, 10, 20, 50, 100, 200
When using the 9602 and 500 A sensor:
10, 20, 50, 100, 200, 500
Example
:CURRENT1:RANGE 50
Sets the current range for channel 1 of
the input unit to 50 A.
Function
Sets the rectifier type (MEAN/RMS) of
current ranging for the specified input
unit
Note
If DC is selected in the COUPling
command, an execution error occurs.
Function
Queries the rectifier type of current
ranging for the specified input unit.
Function
Sets the current ranging of the specified
input unit.
Note
In 1P3W mode or above, for the channel
specification, enter the number of the
lowest-numbered channel in the
combination. For example, when using
input unit channels 1 to 3 in 3V3A
mode, specify "1".
Error
Execution error / If the value which can
not be specified is selected depending on
input unit
:CURRent [
channel no.
] :MEAN
Sets the rectifier type of current ranging.
:CURRent [
channel no.
] :MEAN?
Queries the rectifier type of current ranging.
:CURRent [
channel no.
] :RANGe
Sets the current ranging.
Summary of Contents for Power HiTester 3193
Page 2: ......
Page 50: ...32 3 9 Operations During Power Failure ...
Page 76: ...58 4 13 Degaussing ...
Page 80: ...62 5 2 Setting the Frequency Range fa ...
Page 108: ...90 9 3 Internal Circuit for the External Control and Timing ...
Page 112: ...94 10 3 Output Rate ...
Page 250: ...232 13 9 Error and Overflow Displays ...
Page 278: ...260 17 5 Internal Block Diagram ...
Page 284: ...266 19 2 Installation Procedures For JIS standard For EIA standard External Dimensions ...
Page 300: ...282 20 4 Internal Block Diagram of the 3193 ...
Page 306: ...INDEX 4 Index ...
Page 307: ......
Page 308: ......
Page 309: ......
Page 310: ......