244
────────────────────────────────────────────────────
15.3 Specifications (using with the HIOKI 3193)
────────────────────────────────────────────────────
Voltage (
U
)
Current (
I
)
Active power (
P
)
Power factor influence
(55Hz)
―――――
―――――
±
0.15%f.s.
(power factor=0)
Accuracy of waveform
peak measurement
(voltage or current)
±
1%f.s. (5 Hz to 1 kHz),
±
2%f.s. (1 kHz to 10 kHz),
±
10%f.s. (10 kHz to 100 kHz)
Effective input range: rms value of sine wave is in the effective input range.
Response
(analog output response
time)
FAST (0.1 s): Specification is met for 50 Hz and above.
MID (0.8 s): Specification is met for 10 Hz and above.
SLOW (5.0 s): Specification is met
The analog output response time is the time such that for an input change from
0% to 90% and 100% to 10% of the nominal range value, the value is within
±
1% of the final stabilized value.
LPF switching
OFF/500 Hz (-3dB)
The accuracy specification is met at 60 Hz or less.
phF switching
OFF/200 Hz (-3dB) (Stabilization filter for
U
and
I
polarity determination.)
There is no effect on the
U
,
I
, or
P
accuracy.
Temperature coefficient
(0-18, 28-40℃)
±
0.03%f.s./
℃
or less
Analog output (
U
/
I
/
P
)
(When input DC, sine
wave, range full scale)
DC
±
5 Vf.s.
(1000 V range: DC
±
3.333 V f.s.)
Display accuracy:
±
0.2%f.s
Monitor output (
U
/
I
)
(When input DC, sine
wave, range full scale)
1
Vrms f.s.
(1000 V range: 0.6667 Vrms f.s.
)
Display accuracy:
±
0.2%f.s
Maximum rated voltage
to earth
1000 V
Influence of maximum
input voltage to earth
(1000 Vrms, 50/60 Hz)
±
0.05%f.s. max
(When applying between input terminal and case)
Dielectric strength
(50/60 Hz)
5.55 kV AC
(
1 mA
)
for 1 minute (When using with the 3193)
Between
U
/
I
input terminals and case, between
U
/
I
input terminals and power
plug
Insulation resistance
100 M
Ω
or more at 500 V DC (When using with the 3193)
Between
U
/
I
input terminals and case, between
U
/
I
input terminals and power
plug
Summary of Contents for Power HiTester 3193
Page 2: ......
Page 50: ...32 3 9 Operations During Power Failure ...
Page 76: ...58 4 13 Degaussing ...
Page 80: ...62 5 2 Setting the Frequency Range fa ...
Page 108: ...90 9 3 Internal Circuit for the External Control and Timing ...
Page 112: ...94 10 3 Output Rate ...
Page 250: ...232 13 9 Error and Overflow Displays ...
Page 278: ...260 17 5 Internal Block Diagram ...
Page 284: ...266 19 2 Installation Procedures For JIS standard For EIA standard External Dimensions ...
Page 300: ...282 20 4 Internal Block Diagram of the 3193 ...
Page 306: ...INDEX 4 Index ...
Page 307: ......
Page 308: ......
Page 309: ......
Page 310: ......