4.5 Setting the Scaling (PT/CT/SC Ratios)
────────────────────────────────────────────────────
NOTE
・
For channel combinations of 1P3W and above, the PT ratio, CT ratio, or SC
ratio must be the same for all channels in the combination.
・
When the channel identifies a CT6862, CT ratio of the channel is set to 2.5,
as default. In this state, the current ranges are configured and the measured
values can be read off directly.
Display ranges:
「
1.25(A) / 2.5A / 5A / 12.5(A) / 25A / 50A
」
The default CT ratio is changeable. If you want to change this, please set the
value calculated by the formula: (your desired CT ratio x 2.5). For example,
when you want to set 40 as CT ratio, please set 100(=40 x 2.5).
When CT ratio other than 2.5 (including OFF) is being set, the current ranges
to be displayed are below.
Display ranges:
「
0.5A / 1A / 2A / 5A / 10A / 20A
」
The actual current range configuration with the CT ratio other than 2.5 can
be calculated by multiplying the set CT ratio by each of 0.5A, 1A, 2A, 5A,
10A, 20A.
・
When a CT6865 is used, please set it to CT ratio 2.
The display ranges are [10A/20A/50A/100A/200A/500A], but the range
structure is actually configured to be double the CT ratio after internal
computation.
Summary of Contents for Power HiTester 3193
Page 2: ......
Page 50: ...32 3 9 Operations During Power Failure ...
Page 76: ...58 4 13 Degaussing ...
Page 80: ...62 5 2 Setting the Frequency Range fa ...
Page 108: ...90 9 3 Internal Circuit for the External Control and Timing ...
Page 112: ...94 10 3 Output Rate ...
Page 250: ...232 13 9 Error and Overflow Displays ...
Page 278: ...260 17 5 Internal Block Diagram ...
Page 284: ...266 19 2 Installation Procedures For JIS standard For EIA standard External Dimensions ...
Page 300: ...282 20 4 Internal Block Diagram of the 3193 ...
Page 306: ...INDEX 4 Index ...
Page 307: ......
Page 308: ......
Page 309: ......
Page 310: ......